1.General Introduction
ABB 266GRT gauge pressure transmitter combines with S26 direct or remote seal for gauge pressure measurement, measure range from -40Kpa to 60Mpa, maximum proof pressure is up to 90Mpa and maximum turn down ratio up to 60:1, standard accuracy is 0.04%, meets SIL2,SIL3 safety applications as per IEC 61508 standard.
Depending on actual application, 266GRT allows a gauge pressure measurement using one remote seal directly mounted with a short capillary connecting the integral seal to the transmitter sensor. The S26 Series Seal System is a protective device used to isolate 266GRT gauge pressure transmitter from the process fluid, provides a flexible diaphragm seal between the process fluid and a liquid filled capillary tube connected to the body of the transmitter. The diaphragm isolates the process fluid while the filled capillary tube hydraulically transmits the process pressure to the transmitter sensor. The capillary of remote seal is corrosion-resistant with robust construction in stainless steel with spiral armor protection, also PVC jacket; PVC protection is always recommended except for high temperature application, where stainless steel armor is suggested. The all welded construction assures reliable operation over the widest range of operating temperature and under vacuum conditions.
2.Measuring Span & Range Limits
Sensor Code
|
Lower Range Limit (LRL)
|
Upper Range Limit (URL)
|
Standard TD Ratio
|
Max. TD Ratio
|
Max. Static Pressure
|
C
|
-6Kpa, -60mbar, -24inH2O
|
6Kpa, 60mbar, 24inH2O
|
5:1
|
10:1
|
1Mpa, 10bar, 145psi
|
F
|
-40Kpa, -400mbar, -160inH2O
|
40Kpa, 400mbar, 160inH2O
|
10:1
|
60:1
|
1Mpa, 10bar, 145psi
|
L
|
0.07Kpa, 0.7mbar, -0.5mmHg
|
250Kpa, 2500mbar, 1000inH2O
|
10:1
|
60:1
|
3Mpa, 30bar, 435psi
|
D
|
0.07Kpa, 0.7mbar, -0.5mmHg
|
1000Kpa, 10bar, 145psi
|
10:1
|
60:1
|
6Mpa, 60bar, 870psi
|
U
|
0.07Kpa, 0.7mbar, -0.5mmHg
|
3000Kpa, 30bar, 450psi
|
10:1
|
60:1
|
6Mpa, 60bar, 870psi
|
R
|
0.07Kpa, 0.7mbar, -0.5mmHg
|
10000Kpa, 100bar, 1450psi
|
10:1
|
60:1
|
30Mpa, 300bar, 5945psi
|
V
|
0.07Kpa, 0.7mbar, -0.5mmHg
|
60000Kpa, 600bar, 8700psi
|
10:1
|
60:1
|
90Mpa, 900bar, 13050psi
|
3.Basic Code Selection of 266GRT Gauge Pressure Transmitter
Main Code
|
Suffix Codes
|
Description
|
266GRT
|
|
|
|
|
|
|
|
Gauge pressure transmitter with remote seal, base accuracy 0.04 %
|
SENSOR - Span Limits
|
C
|
|
|
|
|
|
|
0.6 and 6 kPa 6 and 60 mbar 2.41 and 24 in. H2O
|
F
|
|
|
|
|
|
|
0.67 and 40 kPa 6.7 and 400 mbar 2.67 and 160 in. H2O
|
L
|
|
|
|
|
|
|
4.17 and 250 kPa 41.7 and 2500 mbar 16.7 and 1000 in. H2O
|
D
|
|
|
|
|
|
|
16.7 and 1000 kPa 0.167 and 10 bar 2.42 and 145 psi
|
U
|
|
|
|
|
|
|
50 and 3000 kPa 0.5 and 30 bar 7.25 and 435 psi
|
R
|
|
|
|
|
|
|
167 and 10000 kPa 1.67 and 100 bar 24.2 and 1450 psi
|
V
|
|
|
|
|
|
|
1000 and 60000 kPa 10 and 600 bar 145 and 8700 psi
|
Diaphragm Material / Fill Fluid
|
|
R
|
|
|
|
|
|
Diaphragm seal mounted Silicone oil (specify diaphragm seal separately)
|
|
2
|
|
|
|
|
|
Diaphragm seal mounted Fluorocarbon - Galden (specify diaphragm seal separately)
|
|
N
|
|
|
|
|
|
Diaphragm seal mounted White oil (specify diaphragm seal separately)
|
Process Connection Material / Type
|
|
|
R
|
|
|
|
|
Diaphragm seal (except in the case of button diaphragm seals, specify diaphragm seal separately)
|
|
|
G
|
|
|
|
|
Button diaphragm seal (specify button diaphragm seal separately)
|
|
|
M
|
|
|
|
|
Direct mount diaphragm seal (one direct mount seal to be quoted)
|
Housing Material / Electrical Connection
|
|
|
|
A
|
|
|
|
Aluminium alloy (Barrel type) 1/2-14 NPT
|
|
|
|
B
|
|
|
|
Aluminium alloy (Barrel type) M20 x 1.5
|
|
|
|
E
|
|
|
|
Aluminium alloy (Barrel type) Harting Han plug connector (for standard applications)
|
|
|
|
G
|
|
|
|
Aluminium alloy (Barrel type) Fieldbus plug connector (for standard applications)
|
|
|
|
S
|
|
|
|
Stainless steel (Barrel type) 1/2-14 NPT
|
|
|
|
T
|
|
|
|
Stainless steel (Barrel type) M20 x 1.5
|
|
|
|
J
|
|
|
|
Aluminium alloy (DIN type) M20 x 1.5
|
|
|
|
K
|
|
|
|
Aluminium alloy (DIN type) Harting Han plug connector (for standard applications)
|
|
|
|
W
|
|
|
|
Aluminium alloy (DIN type) Fieldbus plug connector (for standard applications)
|
|
|
|
Z
|
|
|
|
Stainless steel (Barrel type) Fieldbus plug connector (for standard applications)
|
Output
|
|
|
|
|
1
|
|
|
HART digital communication and 4 … 20 mA
|
|
|
|
|
2
|
|
|
Profibus PA
|
|
|
|
|
3
|
|
|
Foundation fieldbus
|
|
|
|
|
8
|
|
|
HART digital communication and 4 … 20 mA, SIL2 and SIL3-certified in acc. with IEC 61508
|
Optional Codes
|
/Optional codes
|
4.Basic Code Selection of S26RR Rotating Flange Diaphragm Seals (flush) Ring Joint to ASME B16.5
S26RR
|
|
|
|
|
|
|
|
|
|
|
|
|
Rotating flange diaphragm seal (flush) Ring Joint to ASME B16.5
|
Transmitter Side of Connection
|
H
|
|
|
|
|
|
|
|
|
|
|
|
High pressure side
|
L
|
|
|
|
|
|
|
|
|
|
|
|
Low pressure side
|
Mounting Flange Rating / Size
|
|
D1
|
|
|
|
|
|
|
|
|
|
|
ASME CL150, 1 1/2 inch
|
|
D2
|
|
|
|
|
|
|
|
|
|
|
ASME CL300, 1 1/2 inch
|
|
D3
|
|
|
|
|
|
|
|
|
|
|
ASME CL600, 1 1/2 inch
|
|
D5
|
|
|
|
|
|
|
|
|
|
|
ASME CL900-1500, 1 1/2 inch
|
|
D6
|
|
|
|
|
|
|
|
|
|
|
ASME CL2500, 1 1/2 inch
|
|
E1
|
|
|
|
|
|
|
|
|
|
|
ASME CL150, 2 inch
|
|
E2
|
|
|
|
|
|
|
|
|
|
|
ASME CL300, 2 inch
|
|
E3
|
|
|
|
|
|
|
|
|
|
|
ASME CL600, 2 inch
|
|
E5
|
|
|
|
|
|
|
|
|
|
|
ASME CL900-1500, 2 inch
|
|
E6
|
|
|
|
|
|
|
|
|
|
|
ASME CL2500, 2 inch
|
|
G1
|
|
|
|
|
|
|
|
|
|
|
ASME CL150, 3 inch
|
|
G2
|
|
|
|
|
|
|
|
|
|
|
ASME CL300, 3 inch
|
|
G3
|
|
|
|
|
|
|
|
|
|
|
ASME CL600, 3 inch
|
|
G4
|
|
|
|
|
|
|
|
|
|
|
ASME CL900, 3 inch
|
|
G5
|
|
|
|
|
|
|
|
|
|
|
ASME CL1500, 3 inch
|
|
G6
|
|
|
|
|
|
|
|
|
|
|
ASME CL2500, 3 inch
|
Mounting Flange Material
|
|
|
C
|
|
|
|
|
|
|
|
|
|
Carbon steel
|
|
|
S
|
|
|
|
|
|
|
|
|
|
ANSI 316 SS
|
Extensions Length and Material
|
|
|
|
F
|
|
|
|
|
|
|
|
|
Flush
|
Diaphragm Material
|
|
|
|
|
SM
|
|
|
|
|
|
|
|
ANSI 316 SS, NACE
|
|
|
|
|
HM
|
|
|
|
|
|
|
|
Hastelloy C276, NACE
|
|
|
|
|
LM
|
|
|
|
|
|
|
|
Inconel 625, NACE
|
Seal Surface Finish
|
|
|
|
|
|
3
|
|
|
|
|
|
|
Ring joint
|
Capillary Protection
|
|
|
|
|
|
|
A
|
|
|
|
|
|
ANSI 316L SS armor
|
|
|
|
|
|
|
B
|
|
|
|
|
|
ANSI 316L SS armor with PVC protective cover
|
|
|
|
|
|
|
N
|
|
|
|
|
|
Extension tube for direct mount seal
|
Capillary Length m (Feet)
|
|
|
|
|
|
|
|
1
|
|
|
|
|
Direct-mount construction
|
|
|
|
|
|
|
|
A
|
|
|
|
|
1m, 3 feet
|
|
|
|
|
|
|
|
B
|
|
|
|
|
1.5m, 5 feet
|
|
|
|
|
|
|
|
C
|
|
|
|
|
2m, 7 feet
|
|
|
|
|
|
|
|
D
|
|
|
|
|
2.5m, 8 feet
|
|
|
|
|
|
|
|
E
|
|
|
|
|
3m, 10 feet
|
|
|
|
|
|
|
|
F
|
|
|
|
|
3.5m, 12 feet
|
|
|
|
|
|
|
|
G
|
|
|
|
|
4m, 13 feet
|
|
|
|
|
|
|
|
H
|
|
|
|
|
4.5m, 15 feet
|
|
|
|
|
|
|
|
J
|
|
|
|
|
5m, 17 feet
|
|
|
|
|
|
|
|
K
|
|
|
|
|
5.5m, 18 feet
|
|
|
|
|
|
|
|
L
|
|
|
|
|
6m, 20 feet
|
|
|
|
|
|
|
|
M
|
|
|
|
|
6.5m, 22 feet
|
|
|
|
|
|
|
|
N
|
|
|
|
|
7m, 23.5 feet
|
|
|
|
|
|
|
|
P
|
|
|
|
|
7.5m, 25 feet
|
|
|
|
|
|
|
|
Q
|
|
|
|
|
8m, 27 feet
|
|
|
|
|
|
|
|
R
|
|
|
|
|
9m, 30 feet
|
|
|
|
|
|
|
|
S
|
|
|
|
|
10m, 33 feet
|
|
|
|
|
|
|
|
T
|
|
|
|
|
12m, 40 feet
|
|
|
|
|
|
|
|
U
|
|
|
|
|
14m, 47 feet
|
|
|
|
|
|
|
|
V
|
|
|
|
|
16m, 53 feet
|
Fill Fluid
|
|
|
|
|
|
|
|
|
S
|
|
|
|
Silicon oil PMX 200 10 cst (-40 to 250℃, -40 to 480℉)
|
|
|
|
|
|
|
|
|
P
|
|
|
|
Silicon oil Baysilone PD5 5 cst (-85 to 250℃, -121 to 480℉)
|
|
|
|
|
|
|
|
|
N
|
|
|
|
Inert oil Galden G5, oxygen service
|
|
|
|
|
|
|
|
|
D
|
|
|
|
Inert oil Halocarbon 4.2 oxygen service
|
|
|
|
|
|
|
|
|
G
|
|
|
|
Silicon oil for high temperature (-10 to 375℃, 14 to 707℉)
|
|
|
|
|
|
|
|
|
C
|
|
|
|
Silicon polymer syltherm XLT (-100 to 100℃, -148 to 212℉)
|
|
|
|
|
|
|
|
|
W
|
|
|
|
Mineral oil Esso Marcol 152 (FDA approved)
|
|
|
|
|
|
|
|
|
A
|
|
|
|
Vegetable oil Neobee M-20 (FDA approved)
|
|
|
|
|
|
|
|
|
B
|
|
|
|
Glycerin water 70% (FDA approved)
|
Flushing Ring: Hole and Thread
|
|
|
|
|
|
|
|
|
|
N
|
|
|
None
|
Flushing Ring Material
|
|
|
|
|
|
|
|
|
|
|
N
|
|
None
|
Flushing Ring: Plug and Gasket
|
|
|
|
|
|
|
|
|
|
|
|
N
|
No plug, no gasket
|
Optional Codes
|
/Optional codes
|